{"id":45445,"date":"2022-09-29T10:38:47","date_gmt":"2022-09-29T10:38:47","guid":{"rendered":"https:\/\/biomedpharmajournal.org\/?p=45445"},"modified":"2022-10-01T04:50:10","modified_gmt":"2022-10-01T04:50:10","slug":"the-quality-assurance-for-the-petct-in-nuclear-medicine-evaluation-of-the-daily-quality-control-of-the-positron-emission-tomography","status":"publish","type":"post","link":"https:\/\/biomedpharmajournal.org\/staging\/vol15no3\/the-quality-assurance-for-the-petct-in-nuclear-medicine-evaluation-of-the-daily-quality-control-of-the-positron-emission-tomography\/","title":{"rendered":"The Quality Assurance for the PET\/CT in Nuclear Medicine &#8211; Evaluation of the Daily Quality Control of the Positron Emission Tomography"},"content":{"rendered":"<p><strong>Introduction<\/strong><\/p>\n<p>Generally, clinical radionuclide counting and imaging instruments in nuclear medicine depend on a certain level of accurate and reproducible performance<sup>1<\/sup>. Furthermore, by international standards and recommendations, the quality assurance program should include \u201cQC\u201d procedures. These must be performed on a regular and planned basis, by applicable requirements established by regulatory bodies<sup>2<\/sup>. Indeed \u201cQC\u201d is defined as an essential part of routine nuclear medicine practice. However, over the years many parameters for acceptability testing and performance characterization of the \u201cPET\/CT\u201d, and other nuclear medicine instruments have been developed. To this end, various associations, regulatory, advisory, and professional organizations (American Association of Physicists in Medicine \u201cAAPM\u201d, International Atomic Energy Agency \u201cIAEA\u201d, American Society of Radiology \u201cACR\u201d, and others) have published detailed data collection and analysis methods<sup>1<\/sup>. \u201cPET\/CT\u201d plays an essential role in oncology for diagnosis and follow-up, it is essential to check the equipment regularly, especially for consistency and coincidence, as well as maximum energy through \u201cQC\u201d procedures<sup>3, 4<\/sup>. This technique will continue to play an important role in precision nuclear medicine (4). Indeed, the purpose of \u201cPET\u201d daily quality control is to ensure consistency of detector functionality and to identify any problems, such as a crystal or module failure<sup>5<\/sup>. The procedure is performed using a phantom with a Ge-68 source in a cylindrical form in most cases to ensure optimal crystal performance<sup>3, 5, 6, 7<\/sup>. In addition, the purpose of daily quality control is to correlate the data from the sinograms; sinogram control (uniformity) is obtained daily using a Ge-68 source to the values determined during system calibration, for this the recalibration is required if the differences exceed the tolerance levels<sup>8<\/sup>.The main daily quality control steps related to the operating parameters of the detector are coincidence detection, singles, dead time, timing, and energy resolution<sup>9<\/sup>.<\/p>\n<p>One of the main roles and responsibilities of the qualified nuclear medicine physicist is to ensure optimization of the physical aspects of the diagnostic procedure. This task is performed in collaboration with other qualified medical staff <sup>10<\/sup>. The qualified medical physicist must establish a quality assurance program that will include daily quality control procedures for the equipment in the department that uses ionizing radiation for medical protocols. Personnel of the department should consult with a qualified medical physicist if measured values of daily quality control parameters exceed established tolerances. However, corrective actions should be prescribed to resolve problems raised during daily quality control<sup>11<\/sup>.In addition, a technologist should perform the daily quality control, and it is recommended that these procedures be performed before the clinical use of the scanner<sup>7, 12<\/sup>.<\/p>\n<p><strong>Daily quality control for General Electric Discovery \u201cGE. D\u201d equipment &#8211; \u201cPET\/CT\u201d:<\/strong><\/p>\n<p>An integrated Ge-68 source in the scanner&#8217;s field of view is required for daily quality control of \u201cGE. D\u201d equipment \u201cPET\/CT\u201d; the source is not visible because it is located behind the scanner enclosure<sup>9<\/sup>. As mentioned above, the daily quality control of the \u201cPET\u201d is performed sequentially and evaluates coincidences, singles, timeouts, timing, and system energy. A visual and parametric data report on the status of the detectors in the form of an image \u201csinogram\u201d (Figure.1) after the \u201cPET\u201d daily quality control procedure is completed. A scale of three represents the parametric results obtained during the procedure: green indicates a setting that is within an acceptable range, yellow indicates a recommended setting and red indicates a setting that is outside an acceptable range<sup>13<\/sup>.<strong>\u00a0<\/strong><\/p>\n<table style=\"width: 70%;\" border=\"1\" cellpadding=\"5\">\n<tbody>\n<tr>\n<td><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig1.jpg\"><img decoding=\"async\" class=\"alignnone size-thumbnail wp-image-45457\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig1-150x150.jpg\" alt=\"Vol15No3_The_Red_fig1\" width=\"150\" height=\"150\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig1-150x150.jpg 150w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig1-256x256.jpg 256w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig1.jpg 489w\" sizes=\"(max-width: 150px) 100vw, 150px\" \/><\/a><\/td>\n<td><strong>Figure 1: Example of Sinogram of daily \u201cPET\u201d &#8211; \u201cQC\u201d\u00a0<\/strong><strong>Ge Discovery Technology<sup>13<\/sup><\/strong><strong>\u00a0<\/strong><\/p>\n<p><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig1.jpg\" target=\"_blank\">Click here to view figure<\/a><\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>The following table provides recommendations for the daily quality control procedures that should be scheduled and performed for the \u201cPET\/CT\u201d. The purpose of the procedure, the frequency, and comments on the procedure, the type of technology, and the parameters of \u201cQC\u201d are performed in (Table. 1)<sup>14<\/sup>. In addition, the manufacturer sets the parameters and frequency of \u201cQC\u201d tests for each \u201cPET\/CT\u201d system that each qualified medical physicist should follow to perform a control procedure within the standards<sup>15<\/sup>.<\/p>\n<p><strong>Table 1: Daily quality control for \u201cPET\u201d \u2013 GE systems<sup>14, 15<\/sup><\/strong><\/p>\n<table style=\"width: 95%;\" border=\"1\" cellspacing=\"0\" cellpadding=\"4\">\n<tbody>\n<tr>\n<td style=\"text-align: center;\" width=\"153\"><strong><em>Control<\/em><\/strong><\/td>\n<td style=\"text-align: center;\" width=\"118\"><strong><em>Type of Technology<\/em><\/strong><\/td>\n<td style=\"text-align: center;\" width=\"157\"><strong><em>Purpose<\/em><\/strong><\/td>\n<td style=\"text-align: center;\" width=\"91\"><strong><em>Frequency<\/em><\/strong><\/td>\n<td style=\"text-align: center;\" width=\"169\"><strong><em>Comments<\/em><\/strong><\/td>\n<td style=\"text-align: center;\" width=\"143\"><strong><em>\u00a0<\/em><\/strong><\/p>\n<p><strong><em>QC parameters<\/em><\/strong><\/p>\n<p><strong><em>\u00a0<\/em><\/strong><\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"153\"><strong><em>PET Daily QC<\/em><\/strong><\/td>\n<td style=\"text-align: center;\" width=\"118\">GE Systems<\/td>\n<td style=\"text-align: center;\" width=\"157\">Visualize the operation of the detection modules.<\/p>\n<p>A display of a sinogram<\/td>\n<td style=\"text-align: center;\" width=\"91\">Daily<\/td>\n<td style=\"text-align: center;\" width=\"169\">Performed with point sources (Source of Ge-68)<\/td>\n<td style=\"text-align: center;\" width=\"143\">&nbsp;<\/p>\n<p>Coincidence test<\/p>\n<p>Singles test<\/p>\n<p>Energy test<\/p>\n<p>Timing test<\/p>\n<p>Deadtime test<\/p>\n<p>&nbsp;<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>In effect, the objective of this study is to evaluate the \u201cPET\u201d daily quality control in a nuclear medicine department, of the \u201cPET\u201d to the \u201cPET\/CT\u201d equipment, after one year without a daily control procedure in the absence of a qualified medical physicist in the department.<strong>\u00a0<\/strong><\/p>\n<p><strong>Materials and Methods<\/strong><strong>\u00a0<\/strong><\/p>\n<p>The equipment of this study, especially for the daily quality control has a \u201cPET\/CT\u201d (GE. Discovery 610 Elite, USA), which is composed of a computed tomography part and a \u201cPET\u201d part, using crystals of (Bismuth Germanate &#8211; BGO). A Ge-68 source was used as recommended for this control. Before the start of the \u201cQC\u201d, a questionnaire on the quality assurance program was distributed to the medical staff of the nuclear medicine department to collect responses. Our method is as follows, in the absence of a qualified physicist in the nuclear medicine department, the biomedical engineer of the \u201cPET\/CT\u201d manufacturer started the daily quality control procedure after one year without \u201cQC\u201d, a report was displayed on the console of the \u201cPET\/CT\u201d; sinogram 1; the procedure lasted 11 minutes; (Figure 2). The procedure displayed two errors concerning two modules of the \u201cPET\u201d &#8211; crystal scintillator, the object of our work.<strong>\u00a0<\/strong><\/p>\n<table style=\"width: 70%;\" border=\"1\" cellpadding=\"5\">\n<tbody>\n<tr>\n<td><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig2.jpg\"><img decoding=\"async\" class=\"alignnone size-thumbnail wp-image-45456\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig2-150x150.jpg\" alt=\"Vol15No3_The_Red_fig2\" width=\"150\" height=\"150\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig2-150x150.jpg 150w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig2-256x256.jpg 256w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig2.jpg 487w\" sizes=\"(max-width: 150px) 100vw, 150px\" \/><\/a><\/td>\n<td><strong>Figure 2: Sinogram of daily quality control for the (PET)-<\/strong><strong>GE Discovery 610 Elite (study).<\/strong><\/p>\n<p><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig2.jpg\" target=\"_blank\">Click here to view figure<\/a><\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>Following our method of \u201cQC\u201d, the biomedical engineer launched a roadmap (workflow of the process of calibration of the detector) (Figure 3), to remedy the errors found and perform corrective actions, parameters that refer to the recommendation of the manufacturer. The gain than the energy step (Figure 4) reflects the peak energy spectrum for each PET crystal<sup>16<\/sup>, finally, the timing of each crystal which showed the illustration of a failed module (Figure 5).<\/p>\n<table style=\"width: 70%;\" border=\"1\" cellpadding=\"5\">\n<tbody>\n<tr>\n<td><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig3.jpg\"><img decoding=\"async\" class=\"alignnone size-thumbnail wp-image-45455\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig3-150x150.jpg\" alt=\"Vol15No3_The_Red_fig3\" width=\"150\" height=\"150\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig3-150x150.jpg 150w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig3.jpg 604w\" sizes=\"(max-width: 150px) 100vw, 150px\" \/><\/a><\/td>\n<td><strong>Figure 3: Roadmap (detector calibration process workflow) -GE Discovery 610 Elite (study).<\/strong><\/p>\n<p><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig3.jpg\" target=\"_blank\">Click here to view figure<\/a><\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>&nbsp;<\/p>\n<table style=\"width: 70%;\" border=\"1\" cellpadding=\"5\">\n<tbody>\n<tr>\n<td><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig4.jpg\"><img decoding=\"async\" class=\"alignnone size-thumbnail wp-image-45454\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig4-150x150.jpg\" alt=\"Vol15No3_The_Red_fig4\" width=\"150\" height=\"150\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig4-150x150.jpg 150w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig4-256x256.jpg 256w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig4.jpg 375w\" sizes=\"(max-width: 150px) 100vw, 150px\" \/><\/a><\/td>\n<td><strong>Figure 4: Energy step \u2013Roadmap of the (QC) (PET).<\/strong><\/p>\n<p><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig4.jpg\" target=\"_blank\">Click here to view figure\u00a0<\/a><\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>&nbsp;<\/p>\n<table style=\"width: 70%;\" border=\"1\" cellpadding=\"5\">\n<tbody>\n<tr>\n<td><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig5.jpg\"><img decoding=\"async\" class=\"alignnone size-thumbnail wp-image-45453\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig5-150x150.jpg\" alt=\"Vol15No3_The_Red_fig5\" width=\"150\" height=\"150\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig5-150x150.jpg 150w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig5-256x256.jpg 256w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig5.jpg 457w\" sizes=\"(max-width: 150px) 100vw, 150px\" \/><\/a><\/td>\n<td><strong>Figure 5: Timing step \u2013 Roadmap of the (QC) \u2013 (PET).<\/strong><\/p>\n<p><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig5.jpg\" target=\"_blank\">Click here to view figure<\/a><\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>However, the questionnaire that was distributed to the staff of the nuclear medicine department focused on the following aspects (training in radiation protection for patients and the implementation of the quality assurance program especially the \u201cQC\u201d procedures in the nuclear medicine department.<\/p>\n<p><strong>Results<\/strong><strong>\u00a0<\/strong><\/p>\n<p>At the end of the \u201cQC\u201d procedure, a final report in the form of a sinogram was resulted (Figure 6). Indeed, a positive result, because the failure of the module noted at the beginning of the control, was corrected through the correctives actions performed by the daily quality control procedure especially the steps of the roadmap.<\/p>\n<table style=\"width: 70%;\" border=\"1\" cellpadding=\"5\">\n<tbody>\n<tr>\n<td><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig6.jpg\"><img decoding=\"async\" class=\"alignnone size-thumbnail wp-image-45452\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig6-150x150.jpg\" alt=\"Vol15No3_The_Red_fig6\" width=\"150\" height=\"150\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig6-150x150.jpg 150w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig6-256x256.jpg 256w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig6.jpg 545w\" sizes=\"(max-width: 150px) 100vw, 150px\" \/><\/a><\/td>\n<td><strong>Figure 6: Corrected sinogram &#8211; Daily quality control for the (PET &#8211; GE discovery 610 elite (study).<\/strong><\/p>\n<p><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Red_fig6.jpg\" target=\"_blank\">Click here to view figure<\/a><\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>In addition, according to the results collected from the questionnaire previously distributed to the medical staff of the department of nuclear medicine; object of our work; the staff does not have optimal training in radiation protection for patients in addition to the lack of implementation of a quality assurance program and specifically the establishment of \u201cQC\u201d procedures. The results of the questionnaires distributed reflect in (Table 2).<\/p>\n<table style=\"width: 70%;\" border=\"1\" cellpadding=\"5\">\n<tbody>\n<tr>\n<td><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Reda_tab2.jpg\"><img decoding=\"async\" class=\"alignnone size-thumbnail wp-image-45944\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Reda_tab2-150x150.jpg\" alt=\"Vol15No3_The_Reda_tab2\" width=\"150\" height=\"150\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Reda_tab2-150x150.jpg 150w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Reda_tab2-256x256.jpg 256w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Reda_tab2.jpg 920w\" sizes=\"(max-width: 150px) 100vw, 150px\" \/><\/a><\/td>\n<td><strong>Table 2: Results of questionary (study).<\/strong><\/p>\n<p><a href=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2022\/08\/Vol15No3_The_Reda_tab2.jpg\" target=\"_blank\">Click here to view figure<\/a><\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p><strong>Discussion<\/strong><\/p>\n<p>According to the result of the \u201cQC\u201d, a procedure performed for the \u201cPET\/CT\u201d, object of our this study, in particular the corrective actions of a failing module of the \u201cPET\u201d, it is important to perform the \u201cQC\u201d with the daily frequency for the GE equipment<sup>13,14,15<\/sup>.In addition, according to the results of the questionnaire distributed to the staff of the nuclear medicine department (nuclear medicine physicians, technologists, and nurses), the implementation of a quality assurance program is absent specifically the realization of \u201cQC\u201d procedures, as well as the establishment of training in radiation protection of the patient, particularly continuous training in the department. Two fruitful results of this study demonstrate the importance of a qualified medical physicist within a nuclear medicine department using ionizing radiation. First, he will ensure the implementation of a quality assurance program<sup>11<\/sup>and the establishment of the \u201cQC\u201d procedures that represent an essential step of the said program<sup>2<\/sup>. The procedures will be performed by the \u201cPET\/CT\u201d unit technologist and consulted with a qualified medical physicist<sup>11, 17. <\/sup>Second, he assists in the implementation of training in radiation protection of the patient in the unit, which is also among the steps of the quality assurance program<sup>2<\/sup>.Moreover, it is an important study that affirms the usefulness of the presence of a qualified medical physicist, in particular, a requirement that described the law 142-12 in article 102; at least one qualified medical physicist must be present in each nuclear medicine department<sup>18<\/sup>. Indeed, a study that will strengthen the academic opinion in Morocco in this field.<strong>\u00a0<\/strong><\/p>\n<p><strong>Conclusion<\/strong><strong>\u00a0<\/strong><\/p>\n<p>To conclude this successful study, it is necessary to establish a quality assurance program and implement the culture of radiation safety to ensure patient radiation protection in a nuclear medicine department<sup>2<\/sup>. Indeed, a quality assurance program should be implemented and should include the procedure for daily quality control of \u201cPET\/CT\u201d, including the \u201cPET\u201d part<sup>13<\/sup>.In addition, the presence of a qualified medical physicist becomes essential in the nuclear medicine department, to ensure the establishment of daily quality control procedures and to implement radiation safety recommendations through radiation protection training for the patient. A work that will be in collaboration with the staff of the said department<sup>11<\/sup>.Therefore, it is important to evaluate the daily quality control parameters and take appropriate corrective actions<sup>2 <\/sup>and effective \u201cQC\u201d is an important means of ensuring the repeatability of medical applications<sup>19<\/sup>. To this end, it is made necessary to perform the said in each medical installation through the sources of ionizing radiation<sup>20<\/sup>, thus ensuring optimization of medical exposures and improved image quality for medical protocols in nuclear medicine.<\/p>\n<p><strong>Acknowledgement<\/strong><strong>\u00a0<\/strong><\/p>\n<p>I thank all the contributors who helped me in this work; <em>Khalida Eddaoui<\/em>, <em>Pr. Nouzha Benraiss Aouad<\/em><\/p>\n<p>I am asking for some of your precious time to inform you of the submission of my scientific article, to participate in the development of medical science in general and nuclear medicine in particular.<\/p>\n<p>It will be a great pleasure and a great honor for me to publish in your prestigious journal.<\/p>\n<p>Pending a favorable response, please accept, dear organizers, the expression of my sincere thanks.<\/p>\n<p><strong>Conflict of Interest<\/strong><strong>\u00a0<\/strong><\/p>\n<p>All authors declare that there is no conflict of interest<\/p>\n<p><strong>Funding Source<\/strong><\/p>\n<p>I declare that there are no sources of funding for this research study.<\/p>\n<p><strong>References<\/strong><\/p>\n<ol>\n<li>Kheruka S. C. Importance of quality control in nuclear medicine. International Atomic Energy Agency (IAEA). INIS Volume 52. INIS Issue 23, 2020;<\/li>\n<li>Hallab R, Eddaoui khalida, Ouabi H, Ra\u00efs Aouad NB. Regulatory Requirements of Quality Assurance Program in Nuclear Medicine \u2013 Review of the Procedures. Biomedical and Pharmacology Journal. 2021 Dec 30;14(4):1863\u20137;<br \/>\n<a href=\"https:\/\/doi.org\/10.13005\/bpj\/2284\">CrossRef<\/a><\/li>\n<li>Vyas M, Tafti D. Nuclear Medicine Quality Assurance. [Updated 2022 Feb 5]. In: StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing; 2022 Jan. Available from: http:\/\/https:\/\/www.ncbi.nlm.nih.gov\/books\/NBK576382\/?report=reader.<\/li>\n<li>Schaefferkoetter JD, Osman M, Townsend DW. The Importance of Quality Control for Clinical PET Imaging. Journal of Nuclear Medicine. Technol. 2017 Dec;45(4):265\u20136 ;<br \/>\n<a href=\"https:\/\/doi.org\/10.2967\/jnmt.117.198465\">CrossRef<\/a><\/li>\n<li>Boellaard R, Delgado-Bolton R, Oyen WJG, Giammarile F, Tatsch K, Eschner W, et al. FDG PET\/CT: EANM procedure guidelines for tumor imaging: version 2.0. European Journal of Nuclear Medicine and Molecular Imaging. 2015 Feb;42(2):328\u201354;<br \/>\n<a href=\"https:\/\/doi.org\/10.1007\/s00259-014-2961-x\" target=\"_blank\">CrossRef<\/a><\/li>\n<li>Zanzonico P. Routine Quality Control of Clinical Nuclear Medicine Instrumentation: A Brief Review. Journal of Nuclear Medicine. 2008 Jul;49(7):1114\u201331;<br \/>\n<a href=\"https:\/\/doi.org\/10.2967\/jnumed.107.050203\" target=\"_blank\">CrossRef<\/a><\/li>\n<li>Dondi, M. and all. Quality assurance for PET and PET\/CT systems. Vienna: International Atomic Energy Agency; 2009. (IAEA human health series) ;<\/li>\n<li>Saha GB. Basics of PET Imaging [Internet]. New York, NY: Springer New York; 2010 [cited 2022 Mar 1]. Available from: http:\/\/link.springer.com\/10.1007\/978-1-4419-0805-6;<\/li>\n<li>Claudiu P, Quality Control of Nuclear Medicine Instrumentation and Protocol Standardisation. European Association of Nuclear Medicine.Vienna, Austria, 2017;<\/li>\n<li>Pasawang P, Sontrapornpol T, Krisanachinda A. EXPERIENCE ON PERFORMANCE MEASUREMENTS OF POSITRON EMISSION TOMOGRAPHS, Medical Physics International Journal, vol.7, No.3, 2019;<\/li>\n<li>ACR\u2013AAPM Technical Standard For MEDICAL PHYSICS PERFORMANCE MONITORING OF PET\/CT IMAGING EQUIPMENT, (CSC\/BOC), 2018, The American College of Radiology;<\/li>\n<li>Hac\u0131osmano\u011flu T, Demir M, Toklu T, K\u0131ra\u00e7 FS, \u0130nce M, Parlak Y, and all. Acceptance Tests and Quality Control of the Positron Emission Tomography (PET) Systems. Nuclear Medicine Seminar. 2020 Aug 26;6(2):51\u201370;<br \/>\n<a href=\"https:\/\/doi.org\/10.4274\/nts.galenos.2020.0008\" target=\"_blank\">CrossRef<\/a><\/li>\n<li>Andersen, F. and all. PET\/CT atlas on quality control and image artefacts. \u2014 Vienna: International Atomic Energy Agency, 2014,(IAEA human health series, ISSN 2075\u20133772 ; no. 27) ;<\/li>\n<li>Busemann Sokol E, P\u0142achc\u00ednska A, Britten A, Lyra Georgosopoulou M, et al. Routine quality control recommendations for nuclear medicine instrumentation. European Journal of Nuclear Medicine and Molecular Imaging. 2010 Mar;37(3):662\u201371;<br \/>\n<a href=\"https:\/\/doi.org\/10.1007\/s00259-009-1347-y\" target=\"_blank\">CrossRef<\/a><\/li>\n<li>Lopez BP, Jordan DW, Kemp BJ, Kinahan PE, Schmidtlein CR, Mawlawi OR. PET\/CT acceptance testing and quality assurance: Executive summary of AAPM Task Group 126 Report. Med Phys [Internet]. 2021 Feb [cited 2022 Mar 1];48(2). Available from: https:\/\/onlinelibrary.wiley.com\/doi\/10.1002\/mp.14656;<br \/>\n<a href=\"https:\/\/doi.org\/10.1002\/mp.14656\" target=\"_blank\">CrossRef<\/a><\/li>\n<li>Mawlawi OR, Jordan DW, Halama JR, Schmidtlein CR, Wooten WW. PET\/CT Acceptance Testing and Quality Assurance.The Report of AAPM Task Group 126. 2019 Oct.;<br \/>\n<a href=\"https:\/\/doi.org\/10.37206\/193\" target=\"_blank\">CrossRef<\/a><\/li>\n<li>Hristova I, Boellaard R, Galette P, Shankar LK, Liu Y, Stroobants S, et al. Guidelines for quality control of PET\/CT scans in a multicenter clinical study. European Journal of Nuclear Medicine and Molecular Imaging Physics. 2017 Dec;4(1):23 ;<br \/>\n<a href=\"https:\/\/doi.org\/10.1186\/s40658-017-0190-7\" target=\"_blank\">CrossRef<\/a><\/li>\n<li>Government of Kingdom of Morocco. Law 142-12, Related to Nuclear and Radiological Safety and Security and the Creation of Moroccan Agency for Nuclear and Radiation Safety and Security. 2014;<\/li>\n<li>He Z, Hou L, Sun R, Yin T, Ma P, Chen L, et al. The Status of the Acupuncture Mechanism Study Based on PET\/PET-CT Technique: Design and Quality Control. Evidence-Based Complementary and Alternative Medicine. 2019 Nov 29; 2019:1\u20138;<br \/>\n<a href=\"https:\/\/doi.org\/10.1155\/2019\/9062924\" target=\"_blank\">CrossRef<\/a><\/li>\n<li>Swarna SK, Nivedhitha MS. Awareness of Radiation protection protocols among General practitioners and Endodontists in Chennai-A Cross-sectional questionnaire-based survey. Research Journal of Pharmacy and Technology. 2020;13(1):81.<br \/>\n<a href=\"https:\/\/doi.org\/10.5958\/0974-360X.2020.00015.3\" target=\"_blank\">CrossRef<\/a><\/li>\n<\/ol>\n","protected":false},"excerpt":{"rendered":"<p>Introduction Generally, clinical radionuclide counting and imaging instruments in nuclear  [&#8230;]<\/p>\n","protected":false},"author":15,"featured_media":0,"comment_status":"closed","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"footnotes":""},"categories":[100],"tags":[],"class_list":["post-45445","post","type-post","status-publish","format-standard","hentry","category-vol15no3"],"_links":{"self":[{"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/posts\/45445","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/users\/15"}],"replies":[{"embeddable":true,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/comments?post=45445"}],"version-history":[{"count":5,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/posts\/45445\/revisions"}],"predecessor-version":[{"id":46127,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/posts\/45445\/revisions\/46127"}],"wp:attachment":[{"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/media?parent=45445"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/categories?post=45445"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/tags?post=45445"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}