{"id":1188,"date":"2015-06-18T06:04:28","date_gmt":"2015-06-18T06:04:28","guid":{"rendered":"http:\/\/biomedpharmajournal.org\/?p=1188"},"modified":"2020-04-25T03:08:10","modified_gmt":"2020-04-25T03:08:10","slug":"development-of-draft-design-for-magnet-control-system-of-wireless-endoscopic-capsule","status":"publish","type":"post","link":"https:\/\/biomedpharmajournal.org\/staging\/vol8no1\/development-of-draft-design-for-magnet-control-system-of-wireless-endoscopic-capsule\/","title":{"rendered":"Development of Draft Design for Magnet Control System of Wireless Endoscopic Capsule"},"content":{"rendered":"<p><strong>Introduction<\/strong><\/p>\n<p>At present, a lot of attentionis paid to control systems of wireless endoscopic capsules, because they help to stop the capsule in the exact part of the digestive system for more detailed examination. There are different approaches to control, for example, in [1] expandable capsule is described made of the electroactive polymer, which changes its size when electric current is applied.\u00a0 But control systems are often built on the basis of magnet field usage.<\/p>\n<p>Magnet systems for wireless endoscopic capsules control are described in many patent claims (descriptions) and scientific articles. For example, Sun et al. describe multi-applications of a magnet configuration in actuating wireless capsule endoscope [2]. Lien et al. propose a capsule endoscope magnetic control system [3]. Wakefield tells about magnetically propelled capsule endoscopy provides for the medical examination and treatment of the gastrointestinal tract, reproductive tract, trachea, lungs, vascular system or any accessible body cavity [4]. Kim et al. present a modified magnetic capsule endoscope that can be fixed inside the stomach and to monitor the gastric motility [5].<\/p>\n<p>The described methods differ from the one, described in this article. The developed solution helps to provide more accurate and quick magnet field change (in this case magnetic field distribution can be stable and can guarantee, that the capsule will move in the required direction and with the necessary speed), system fault tolerance increases.<\/p>\n<p>Control system development should be based on the results of detailed numerical modeling and prototype creation, which helps to check modeling results, to work in practice developed approaches to issue resolution and form general technical specifications for the system.<\/p>\n<p><strong>Materials and Methods<\/strong><\/p>\n<p>To evaluate sample coil parameters it is necessary to estimate, which fields and field gradients are required for creation of efforts and moments, which can move and spin the coil.For this purpose preliminary analytical estimations were made. To solve the issue at first it is necessary to make some resistance to motion of the capsule to simulate the movement within intestine. At second, on the first stage it is convenient to compensate gravity to take into account only the interaction between the magnet field and the environmental resistance. That is why on this stage it is more preferable to put the capsule with some thick liquid, like glycerin. Thus, to evaluate the necessary power for capsule movement it was considered that is can be compared with thick friction force within the liquid [6]:<\/p>\n<p>F<sub>fr<\/sub>= 3\u03c0d\u03bd\u03b7 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 (1)<\/p>\n<p>where d is a capsule diameter equal to 0.012 m, movement speed (\u03bd) equals10 mm\/sec (this speed is enough to move through 4-5 m tract for a reasonable amount of time), \u03b7 \u2013 dynamical liquid viscosity.<\/p>\n<p>Means, for glycerin with viscosity 1480* Pa*secat a temperature of 20\u00b0\u0421 the friction force will be:<\/p>\n<p>F<sub>fr<\/sub>= 3*3.14*0.012(m)*0.01(m\/sec)*1480* (Pa*sec) = 1.67* \u041d = 0.167 g.wt<\/p>\n<p>Let us assume that some threshold value of force is equal to 0.2 g.wt.To determine the limit spin moment produced by the magnet field, let us evaluate the moment of a couple for friction force, evaluate the moment of a couple for friction forceaffecting the capsule, which is spinning around the axis perpendicular to the symmetry axis:<\/p>\n<p>N<sub>fr<\/sub>= 2*F<sub>fr<\/sub>*L\/2=3\u03c0d\u03bd\u03b7*L\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0 (2)<\/p>\n<p>where L is the length of the capsule, thus:<\/p>\n<p>N<sub>fr<\/sub> = 0.167(g.wt)*3(sm) = 0.5 g.wt*sm.<\/p>\n<p>Now for the evaluated limit values of force and moment let us calculate the coils parameters. Consider that the working volume, where the capsule will be places, is a cube 200 mm on a side. Because it was necessary to preliminary evaluate the coils parameters, we assume a coil with one circular turn with the current.<\/p>\n<p>Field of circular turn axis depends on the distance from the turn center [7]:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1194\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f2.jpg\" alt=\"Formula\" width=\"603\" height=\"142\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f2-300x71.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f2.jpg 603w\" sizes=\"(max-width: 603px) 100vw, 603px\" \/><\/p>\n<p>&nbsp;<\/p>\n<p>Let us make a diagram of this dependence for the case, when the coil is recorded. In Fig. 1 the dependence of coefficient before\u00a0\u03bc<sub>0<\/sub>I\/R<sup>2\u00a0<\/sup>from the distance along Z axis between the coils in coil radius units; one turn is powered with current and is situated in -1 coordinate of coil radius.<\/p>\n<table style=\"width: 70%;\" border=\"1\" cellpadding=\"5\">\n<tbody>\n<tr>\n<td style=\"text-align: center;\">\u00a0<img decoding=\"async\" class=\"alignnone size-thumbnail wp-image-1217\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig1-150x150.jpg\" alt=\"Figure 1\" width=\"150\" height=\"150\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig1-150x150.jpg 150w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig1-256x256.jpg 256w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig1.jpg 473w\" sizes=\"(max-width: 150px) 100vw, 150px\" \/><\/td>\n<td>\u00a0<strong>Figure 1<\/strong><\/p>\n<p><a href=\"http:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig1.jpg\" target=\"_blank\">Click here to View figure<\/a><\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>&nbsp;<\/p>\n<p><sup>To determine power characteristics of the coil it is necessary to take into consideration the least force influencing a dipole. This force is influencing the dipole placed in the furthest dot from the \u201cpulling\u201d coil with the least gradient on the distance of two coil radiuses. Zero gradient in the center of the turn was not taken into consideration, because this dot is knowingly placed outside of the working area.<\/sup><\/p>\n<p>Further magnetic moment for the magnet was determined. The magnet placed inside the capsule will be considered in approximation of the magnet dipole with the magnet moment and will be calculated according to the formula:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1196\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f5.jpg\" alt=\"Formula\" width=\"365\" height=\"52\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f5-300x43.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f5.jpg 365w\" sizes=\"(max-width: 365px) 100vw, 365px\" \/><\/p>\n<p>where\u00a0<em>Br<\/em> is a remanent magnetic induction of the N38 magnet equal to 1.25 T, and V is the magnet\u2019s volume, therefore the magnet moment equals:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1198\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f5B.jpg\" alt=\"Formula\" width=\"300\" height=\"52\" \/><\/p>\n<p>Let us put the force expression, which affects the dipole and is situated on the symmetry axis in the center of the system and in the center of the opposite turn without power (z = 2R) depending on the number of ampere-turns and turn radius:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1199\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f67.jpg\" alt=\"Formula\" width=\"649\" height=\"130\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f67-300x60.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f67.jpg 649w\" sizes=\"(max-width: 649px) 100vw, 649px\" \/><\/p>\n<p>&nbsp;<\/p>\n<p><strong>Results<\/strong><\/p>\n<p>Consider next several coil systems: two systems made from the copper hollow tube 6.35 mm in diameter, two systems made from copper wires 1 mm in diameter and two systems made from the copper wire 2 mm in diameter.<\/p>\n<ol>\n<li>Copper hollow tube. Characteristics:<\/li>\n<\/ol>\n<ul>\n<li>Coil cross-section 5 smx 10 sm.<\/li>\n<li>Innercoildiameter \u2013 20<\/li>\n<li>Outercoildiameter \u2013 30 sm.<\/li>\n<li>Mid diameter\u2013 25 sm.<\/li>\n<li>Diameter \u2013 3\/8 inches or 6.35 mm in diameter.<\/li>\n<li>Thickness \u2013 0.76<\/li>\n<li>Wire section\u2013 7.12mm<sup>2<\/sup>.<\/li>\n<\/ul>\n<p>Horizontal slice of coil system with 5sm x 10 smsection is presented in Fig. 2.<\/p>\n<p>Depending on the gap between the tubes, the following numbers of turns can be evaluated with 10 x 5 smcross-section (Table 1):<\/p>\n<p>Without additional cooling the 100 A current can flow through such tube. For example, if 100 A current is used, than with the most density packaging of 100 turns, 10kA-turns can be taken as well as force on the boarder:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1200\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7B.jpg\" alt=\"Formula7\" width=\"510\" height=\"229\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7B-300x135.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7B.jpg 510w\" sizes=\"(max-width: 510px) 100vw, 510px\" \/><\/p>\n<p>Current source strength:<\/p>\n<p><em>P = I<sup>2<\/sup>R = \u00a0(100A)<sup>2<\/sup>* 0.20 hm=2000Wt<\/em><\/p>\n<p>Power voltage:<\/p>\n<p>U = IR \u00a0= 100 \u0410* \u00a0= 20 W<\/p>\n<p>2. \u00a0 \u00a0Copper tube. Characteristics:<\/p>\n<ul>\n<li>Coil cross-section 10sm x 10 sm.<\/li>\n<li>Coil made with copper tubes.<\/li>\n<li>Inner coil diameter \u2013 20<\/li>\n<li>Outercoildiameter \u2013 30<\/li>\n<li>Diameter \u2013 3\/8 inches or 6.35 mm in diameter.<\/li>\n<li>Thickness \u2013 0.76<\/li>\n<li>Cross section \u2013 7.12mm<sup>2<\/sup>.<\/li>\n<li>Middiameter\u2013 30<\/li>\n<\/ul>\n<p>Horizontal slice of coil system with 10 sm x 10 sm section is presented in Fig. 3.<\/p>\n<p>Depending on the gap between the tubes, the following numbers of turns can be evaluated with 10 x10 sm cross-section (Table 2):<\/p>\n<p>For the same 100\u0410and coil radius equal to 15 sm the following force will be on the boarder:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1201\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7C.jpg\" alt=\"Formula\" width=\"392\" height=\"220\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7C-300x168.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7C.jpg 392w\" sizes=\"(max-width: 392px) 100vw, 392px\" \/><\/p>\n<p>&nbsp;<\/p>\n<p>Current source strength:<\/p>\n<p><em>P = I<sup>2<\/sup>R = \u00a0(100A)<sup>2<\/sup>* 0.50 hm=5000 Wt<\/em><\/p>\n<p>Power voltage:<\/p>\n<p><em>U = IR \u00a0= 100 \u0410* 0.50hm \u00a0= 50 W<\/em><\/p>\n<p>In comparison to the previous case, it is clear that we need twice the amount of copperwire and the force has raised only by 20%.<\/p>\n<p>3. \u00a0 \u00a0Copper wire with 1 mm in diameter.Characteristics:<\/p>\n<ul>\n<li>Coil cross-section 5smx 10 s<\/li>\n<li>Inner coil diameter \u2013 20<\/li>\n<li>Outercoildiameter \u2013 30<\/li>\n<li>Mid diameter \u2013 25<\/li>\n<li>Wire diameter \u2013 1<\/li>\n<li>Wire section = 0.785mm<sup>2<\/sup>.<\/li>\n<li>Number of turns = 50*100 = 5000 turns.<\/li>\n<\/ul>\n<p>Horizontal slice of coil system is presented in Fig. 2.<\/p>\n<p>If current density is equal to 1\u0410\/mm<sup>2<\/sup>, means this density will be taken for evaluation of the corresponding transformer coils, so the current equal to 0.8A will flow; force at the boarders of the working volume will be equal to:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1205\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7D.jpg\" alt=\"Vol8_No1_Deve_Mikh_f7D\" width=\"505\" height=\"228\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7D-300x135.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7D.jpg 505w\" sizes=\"(max-width: 505px) 100vw, 505px\" \/><\/p>\n<p>&nbsp;<\/p>\n<p>Current source strength:<\/p>\n<p><em>P = I<sup>2<\/sup>R = \u00a0(0.8A)<sup>2<\/sup>* 900\u00a0m=57.6 Wt<\/em><\/p>\n<p>Power voltage:<\/p>\n<p>U = IR \u00a0= 0.8 \u0410* 900m \u00a0= 72 W<\/p>\n<p>4. \u00a0 \u00a0 Copper wire with 1 mm in diameter. Characteristics:<\/p>\n<ul>\n<li>Coil cross-section 10sm x 10 sm.<\/li>\n<li>Inner coil diameter \u2013 20<\/li>\n<li>Outercoildiameter \u2013 40<\/li>\n<li>Mid diameter\u2013 30<\/li>\n<li>Wire diameter\u2013 1<\/li>\n<li>Wire section= 0.785mm<sup>2<\/sup>.<\/li>\n<li>Number of turns= 100*100 = 10000<\/li>\n<\/ul>\n<p>Horizontal slice of coil system is presented in Fig. 3.<\/p>\n<p>For the currency density equal to 1\u0410\/mm<sup>2<\/sup>the voltage will be 0.8\u0410;force at the boarders of the working volume will be equal to:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1206\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7E.jpg\" alt=\"Formula\" width=\"514\" height=\"230\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7E-300x134.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7E.jpg 514w\" sizes=\"(max-width: 514px) 100vw, 514px\" \/><\/p>\n<p>&nbsp;<\/p>\n<p>Current source strength:<\/p>\n<p><em>P = I<sup>2<\/sup>R = \u00a0(0.8A)<sup>2<\/sup>* 2160 hm=\u00a0138Wt<\/em><\/p>\n<p>Power voltage:<\/p>\n<p><em>U = IR \u00a0= 0.8 \u0410* 21260 hm \u00a0= 173 W<\/em><\/p>\n<p>5. \u00a0 Copper wire with 2 mm in diameter. Characteristics:<\/p>\n<ul>\n<li>Coil 5sm x 10 cm cross-section.<\/li>\n<li>Inner coil diameter \u2013 20 sm.<\/li>\n<li>Outer coil diameter \u2013 30 sm.<\/li>\n<li>Mid diameter \u2013 25 sm.<\/li>\n<li>Wire diameter \u2013 2 mm.<\/li>\n<li>Wire section = 3.14 mm<sup>2<\/sup>.<\/li>\n<li>Number of turns= 25*50 = 1250 turns.<\/li>\n<\/ul>\n<p>Horizontal slice of coil system is presented in Fig. 2.<\/p>\n<table style=\"width: 70%;\" border=\"1\" cellpadding=\"5\">\n<tbody>\n<tr>\n<td style=\"text-align: center;\">\u00a0<img decoding=\"async\" class=\"alignnone size-thumbnail wp-image-1218\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig2-150x150.jpg\" alt=\"Figure 2\" width=\"150\" height=\"150\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig2-150x150.jpg 150w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig2-256x256.jpg 256w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig2.jpg 514w\" sizes=\"(max-width: 150px) 100vw, 150px\" \/><\/td>\n<td><strong>\u00a0Figure 2<\/strong><\/p>\n<p><a href=\"http:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig2.jpg\" target=\"_blank\">Click here to View figure<\/a><\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>&nbsp;<\/p>\n<p>For the currency density equal to 1\u0410\/mm<sup>2<\/sup> the voltage will be 3.14\u0410;force at the boarders of the working volume will be equal to:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1209\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7F.jpg\" alt=\"Formula\" width=\"526\" height=\"230\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7F-300x131.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7F.jpg 526w\" sizes=\"(max-width: 526px) 100vw, 526px\" \/><\/p>\n<p>Current source strength:<\/p>\n<p><em>P = I<sup>2<\/sup>R = \u00a0(3.14A)<sup>2<\/sup>* 5.\u00a060 hm= 55Wt<\/em><\/p>\n<p>Power voltage:<\/p>\n<p><em>U = IR \u00a0= 3.14 A* 5.60 hm \u00a0= 18 W<\/em><\/p>\n<p>6. \u00a0 \u00a0 \u00a0 Copper wire with 2 mm in diameter. Characteristics:<\/p>\n<ul>\n<li>Coil cross-section 10sm x 10 sm.<\/li>\n<li>Inner coil diameter \u2013 20<\/li>\n<li>Outercoildiameter \u2013 40<\/li>\n<li>Mid diameter\u2013 30<\/li>\n<li>Wire diameter \u2013 2<\/li>\n<li>Wire section= 3.14 mm<sup>2<\/sup>.<\/li>\n<li>Number of turns= 500*500 = 2500 turns.<\/li>\n<\/ul>\n<p>Horizontal slice of coil system is presented in Fig 3.<\/p>\n<table style=\"width: 70%;\" border=\"1\" cellpadding=\"5\">\n<tbody>\n<tr>\n<td style=\"text-align: center;\">\u00a0<img decoding=\"async\" class=\"alignnone size-thumbnail wp-image-1219\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig3-150x150.jpg\" alt=\"Figure 3\" width=\"150\" height=\"150\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig3-150x150.jpg 150w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig3-256x256.jpg 256w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig3.jpg 540w\" sizes=\"(max-width: 150px) 100vw, 150px\" \/><\/td>\n<td><strong>\u00a0Figure\u00a03<\/strong><\/p>\n<p><a href=\"http:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_fig3.jpg\" target=\"_blank\">Click here to View figure<\/a><\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>&nbsp;<\/p>\n<p>For the currency density equal to 1\u0410\/mm<sup>2<\/sup> the voltage will be 3.15\u0410;force at the boarders of the working volume will be equal to:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1210\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7G.jpg\" alt=\"Formula\" width=\"522\" height=\"230\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7G-300x132.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f7G.jpg 522w\" sizes=\"(max-width: 522px) 100vw, 522px\" \/><\/p>\n<p>&nbsp;<\/p>\n<p>Current source strength:<\/p>\n<p><em>P = I<sup>2<\/sup>R = \u00a0(3.14A)<sup>2<\/sup>* 13.5 Ohm= 133\u00a0Wt<\/em><\/p>\n<p>Power voltage:<\/p>\n<p><em>U = IR \u00a0= 3.14 A* 13.5\u00a0Ohm \u00a0= 42\u00a0W<\/em><\/p>\n<p>7. \u00a0 Basic variant.<\/p>\n<p>The coil is made of several single coils with air-gap between them for the possibility of airing. (Fig.4).<\/p>\n<p>For a wire with 1 mm in diameter single coil parameters are:<\/p>\n<ul>\n<li>Wire diameter\u2013 1<\/li>\n<li>Inner diameter \u2013 20<\/li>\n<li>Outerdiameter \u2013 30<\/li>\n<li>Cross-section \u2013 5sm x 2 sm.<\/li>\n<li>Wire section \u2013 0.785 mm<sup>2<\/sup>.<\/li>\n<li>Number of turns \u2013 1000.<\/li>\n<\/ul>\n<p>Thus, the resistance of this coil is:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1212\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f8A.jpg\" alt=\"Formula\" width=\"405\" height=\"56\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f8A-300x41.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f8A.jpg 405w\" sizes=\"(max-width: 405px) 100vw, 405px\" \/><\/p>\n<p>&nbsp;<\/p>\n<p>For 10 such coils the power equals:<\/p>\n<p>P = 10 P<sub>0<\/sub> = 720 Wt<\/p>\n<p>This power can be provided by the \u0410\u041a\u0418\u041f-1134-300-5 power sources (Capacity 1.5 kW, Voltage 0-300W, Current 0-5\u0410).<\/p>\n<p>Single coil wire length:<\/p>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1215\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f9.jpg\" alt=\"Formula\" width=\"420\" height=\"114\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f9-300x81.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f9.jpg 420w\" sizes=\"(max-width: 420px) 100vw, 420px\" \/><\/p>\n<p>For a wire 2mm in diameter:<\/p>\n<p>Single coil parameters:<\/p>\n<ul>\n<li>Wire diameter\u2013 2<\/li>\n<li>Inner diameter \u2013 20<\/li>\n<li>Outer diameter \u2013 30<\/li>\n<li>Cross-section \u2013 5sm x 2 sm.<\/li>\n<li>Wire section \u2013 3.14 mm<sup>2<\/sup>.<\/li>\n<li>Number of turns \u2013 250.<\/li>\n<li>Resistance of such coil equals:<\/li>\n<\/ul>\n<p><img decoding=\"async\" class=\"alignnone size-full wp-image-1216\" src=\"https:\/\/biomedpharmajournal.org\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f10.jpg\" alt=\"Formula\" width=\"403\" height=\"61\" srcset=\"https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f10-300x45.jpg 300w, https:\/\/biomedpharmajournal.org\/staging\/wp-content\/uploads\/2015\/11\/Vol8_No1_Deve_Mikh_f10.jpg 403w\" sizes=\"(max-width: 403px) 100vw, 403px\" \/><\/p>\n<p>&nbsp;<\/p>\n<p>With current equal to 8\u0410:<\/p>\n<p><em>P<sub>0<\/sub> = I<sup>2<\/sup>R = (8A)<sup>2<\/sup> 1.13 Ohm = 72 Wt<\/em><\/p>\n<p>For 10 such coils, the capacity is: P= 10 P<sub>0<\/sub> = 720 Wt<\/p>\n<p>Single coil wire length:<\/p>\n<p><em>L = 2*\u03c0* 0.125 (m)* 250 (turns) = 196m<\/em><\/p>\n<p>The weight of the wire of such length:<\/p>\n<p>M = plS = 8900 (kg\/m<sup>3<\/sup>)* 196(m)* 3.14* 10<sup>-6<\/sup>(m<sup>2<\/sup>)=5.5 kg<\/p>\n<p><strong>Table 1: Gaps and number of turns<\/strong><\/p>\n<table style=\"width: 95%;\" border=\"1\" cellspacing=\"0\" cellpadding=\"4\">\n<tbody>\n<tr>\n<td style=\"text-align: center;\" width=\"55%\"><strong>Gap between the turns <\/strong><\/td>\n<td style=\"text-align: center;\" width=\"44%\"><strong>Number of turns<\/strong><\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"55%\">1<\/td>\n<td style=\"text-align: center;\" width=\"44%\">98<\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"55%\">2<\/td>\n<td style=\"text-align: center;\" width=\"44%\">72<\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"55%\">3<\/td>\n<td style=\"text-align: center;\" width=\"44%\">55<\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"55%\">4<\/td>\n<td style=\"text-align: center;\" width=\"44%\">50<\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"55%\">5<\/td>\n<td style=\"text-align: center;\" width=\"44%\">45<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>&nbsp;<\/p>\n<p><strong>Table 2: Gaps and number of turns<\/strong><\/p>\n<table style=\"width: 95%;\" border=\"1\" cellspacing=\"0\" cellpadding=\"4\">\n<tbody>\n<tr>\n<td style=\"text-align: center;\" width=\"54%\"><strong>Gap between the turns <\/strong><\/td>\n<td style=\"text-align: center;\" width=\"45%\"><strong>Number of turns<\/strong><\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"54%\">1<\/td>\n<td style=\"text-align: center;\" width=\"45%\">196<\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"54%\">2<\/td>\n<td style=\"text-align: center;\" width=\"45%\">144<\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"54%\">3<\/td>\n<td style=\"text-align: center;\" width=\"45%\">110<\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"54%\">4<\/td>\n<td style=\"text-align: center;\" width=\"45%\">100<\/td>\n<\/tr>\n<tr>\n<td style=\"text-align: center;\" width=\"54%\">5<\/td>\n<td style=\"text-align: center;\" width=\"45%\">90<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>&nbsp;<\/p>\n<p><strong>Conclusions<\/strong><\/p>\n<p>According to the above mentioned calculations for sample coil, the following conclusions can be made:<\/p>\n<ol>\n<li>First, two different geometry for coil cross-section, square one (10 sm on side) andrectangular one (5 sm and 10 sm on sides) produce almost equal forces, which affect dipole (almost 25%), but rectangular section will have better heat characteristics, because it has more surface area-inner volume ratio. Surface area provides cooling and within the inner volume the heating is made.<\/li>\n<li>Second, the power sources used (\u0410\u041a\u0418\u041f-1134-300-5) help to work with high voltage and with low currents that is why it is more preferable to use copper wire, than copper tube.But the proposed calculation shows that with equal parameters, tubes coils provide more force, means when copper tubes are used the system is more compact.<\/li>\n<li>The proposed calculations help to choose optimal coil construction for magnetic control system for wireless endoscopic capsule. They can be used for improving technical parameters of the control system in design.<\/li>\n<\/ol>\n<p>The authors plan to make a sample of the chosen magnet coil and to test it.<\/p>\n<p><strong>Acknowledgement<\/strong><\/p>\n<p>The publication is prepared in accordance with the scientific research under the Agreement between \u201cMobile informatics\u201d (LLC) and Ministry of Education and Science of the Russian Federation No. 14.579.21.0053 dated 23.09.2014. Unique project identification number is RFMEFI57914X0053.<\/p>\n<p><strong>References<\/strong><\/p>\n<ol>\n<li>Dmitry Mikhaylov, TimurKhabibullin, Igor Zhukov, Andrey Starikovskiy, LandishGubaydulina, Natalya Romanchuk and Vladimir Konev. Development of Retention System of the Autonomous Endoscopic Capsule and Its Functionalities. BIODEVICES 2014. Proceedings of the International Conference on Biomedical Electronics and Devices. ESEO, Angers, Loire Valley, France. 3 &#8211; 6 March, 2014. Pages 77-84.<\/li>\n<li>Sun, Z.-J., Cheng, X.-G., Cao, S., Ye, B., Zhang, H.-H., Liu, S. Multi-applications of a magnet configuration in actuating capsule endoscope. IEEE\/ASME International Conference on Advanced Intelligent Mechatronics, 2014. Pages 106-111.<\/li>\n<li>Gi-Shih Lien, Chih-Wen Liu, Joe-Air Jiang, Cheng-Long Chuang, Yu-Hao Chang, Wen-Chi Huang. Capsule endoscope magnetic control system. Patent application: US 20150087898 A1, publication date Mar 26, 2015.<\/li>\n<li>Wakefield Glenn. Magnetically propelled capsule endoscopy. Patent application: WO 2004086958 A1, publication date Oct 14, 2004.<\/li>\n<li>Kim, H.M., Choi, J.S., Cho, J.H. A pilot trial of ambulatory monitoring of gastric motility using a modified magnetic capsule endoscope. Journal of Neurogastroenterology and Motility, Volume 20, Issue 2, 2014, Pages 261-264.<\/li>\n<li>G.Fluid mechanics, 3 ed., \u041c., 1970.<\/li>\n<li>Slobodanyuk\u0410.I. Physics 10. \u00a712. Permanent magnet field. URL: http:\/\/www.physbook.ru\/index.php\/\u0421\u043b\u043e\u0431\u043e\u0434\u044f\u043d\u044e\u043a_\u0410.\u0418._\u0424\u0438\u0437\u0438\u043a\u0430_10\/12.7.<\/li>\n<\/ol>\n","protected":false},"excerpt":{"rendered":"<p>Introduction At present, a lot of attentionis paid to control  [&#8230;]<\/p>\n","protected":false},"author":4,"featured_media":0,"comment_status":"closed","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":{"footnotes":""},"categories":[3],"tags":[],"class_list":["post-1188","post","type-post","status-publish","format-standard","hentry","category-vol8no1"],"_links":{"self":[{"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/posts\/1188","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/users\/4"}],"replies":[{"embeddable":true,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/comments?post=1188"}],"version-history":[{"count":4,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/posts\/1188\/revisions"}],"predecessor-version":[{"id":32837,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/posts\/1188\/revisions\/32837"}],"wp:attachment":[{"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/media?parent=1188"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/categories?post=1188"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/biomedpharmajournal.org\/staging\/wp-json\/wp\/v2\/tags?post=1188"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}