Mohammed D. H, Fatalla A. A, Jani G. H. Comparison of Some Mechanical and Physical Properties of Three Types of Impression Materials with Different Dental Implant Angulations. Biomed Pharmacol J 2018;11(3).
Manuscript received on :10-Jun-2018
Manuscript accepted on :27-July-2018
Published online on: 29-08-2018
Plagiarism Check: Yes
Reviewed by: Joao Paulo Schwartz
Second Review by: Ayush Goyal
Final Approval by: Dr. Javad Sharifi-Rad

How to Cite    |   Publication History
Views Views: (Visited 2,403 times, 1 visits today)   Downloads PDF Downloads: 685

Dhuha H. Mohammed, Abdalbseet A. Fatalla and Ghassak H. Jani

Department of Prosthodontic, College of Dentistry, University of Baghdad, Iraq.

Corresponding Author E-mail: abdalbasit@codental.uobaghdad.edu.iq

DOI : https://dx.doi.org/10.13005/bpj/1499

Abstract

Choosing an appropriate impression material is a challenge for many dentists, yet an essential component to provide an excellent clinical outcome and improve productivity and profit. The purpose of present study was to compare wettability, tear strength and dimensional accuracy of three elastomeric impression materials, with the same consistencies (light-body). Three commercially available light body consistency and regular set 3M ESPE Express polyvinylsiloxane (PVS), 3M ESPE Permadyne polyether (PE), and Identium (ID), impression materials were comparedTear strength test, contact angle test and linear dimensional accuracy were evaluated for three elastic impression material. Among the three experimental groups PE impression material exhibited the higher mean values of tear strength, followed by ID group then PVS which showed the lowest mean value. For wettability test, Polyether group exhibited the lowest mean values of contact angle, followed by mean values of contact angle of ID group, while the PVS group showed the highest contact angle. There was a significant difference in the linear distance measurements between the two parallel impression coping/analogue assembly of the working models obtained by using (Identium, 3M ESPE ExpressPVS , 3M ESPE Permadyne polyether) the polyether impression materials provide higher tear strengths and lower wettability than elastomer impression materials and Identium material have acceptable tear strengths and wettability. All three experimental impression materials were distorted in both conditions (angled and parallel implant situations)

Keywords

Linear dimensional accuracy;  Polyether; Polyvinylsiloxane; Tear strength

Download this article as: 
Copy the following to cite this article:

Mohammed D. H, Fatalla A. A, Jani G. H. Comparison of Some Mechanical and Physical Properties of Three Types of Impression Materials with Different Dental Implant Angulations. Biomed Pharmacol J 2018;11(3).

Copy the following to cite this URL:

Mohammed D. H, Fatalla A. A, Jani G. H. Comparison of Some Mechanical and Physical Properties of Three Types of Impression Materials with Different Dental Implant Angulations. Biomed Pharmacol J 2018;11(3). Available from: http://biomedpharmajournal.org/?p=21902

Introduction

Choosing an appropriate impression material is a challenge for many dentists, yet an essential component to providing an excellent clinical outcome and improving productivity and profit. However, with the wide array of impression materials available, it is often difficult to choose the proper product for each situation. the accuracy of an impression rely on many factors , so choosing appropriate impression technique combine with the proper impression material reducedthe requiring adjustment and lead to a well–fitting restoration.1-3

Moist environment nature of the mouth suggests that itshould bedried with air syringes, anti–sialogogues, cotton rolls, and dry pads cause of saliva are often present along with Crevicular fluid and blood even with the best retraction techniques.

When it comes to precision , the most common elastomeric impression materials currently used for making fixed and removable prosthodontic restorations are polyvinylsiloxane (PVS) and polyether (PE).2,4

On one hand, in many situations, polyether provide consistent results. It also can provide good accuracy and surface detail as well as low shrinkage upon setting. Dimensional stability superior to their counterparts PVS.5,6

Thus even though these decent properties ,it characterized bydifficulty of removing impressions made of polyether from the mouth, and also an increased risk of die breakage, could be associated with the lower flexibility of these materials and higher stiffness when compared to other elastomeric materials.6

On the other hand, polyvinylsiloxane are accurate impression materials with excellent dimensional stability, good detail reproduction, high tear strength, adequate working time, and high recovery from deformation. Although meeting many of the criteria for an ideal impression material, polyvinylsiloxanesintrinsically are hydrophobic in nature, which can result in voids at the margin of the tooth preparation in the impression and bubbles in gypsum casts. However, VPS materials are recently being labeled as hydrophilic due to the addition of extrinsic surfactants.7-9

The latest class of impression material is the vinyl-polyether hybrids that include identium..Furthermore, in 2009 presented a newly impression material called vinylsiloxanether (PVSE) (Identium, Kettenbach Company, Eschenburg, Germany). It combined chemically a polyether material and a polyvinylsiloxane, this commercially produced impression material are theoretically claimed to purchaseelements, properties, and benefits of both impression materials VPS and PE

PVES is supplied as a 2-paste auto mixing system and contains a polymer with polyether and siloxane (e.g., addition silicone) groups that are promoted as hydrophilic materials that presumably maintain the stability and characteristics of the parent products.10,11

The purpose of this study was to compare tear strength,wettability and dimensional accuracy of three elastomeric impression materials,with the same consistencies (light-body).

The null hypothesis stated that there would be no significant difference in tear strength,wettability and dimensional accuracy among polyether vinylsiloxanether andpolyvinysiloxane impression.

Material and Methods

Three commercially available light body consistency and regular set 3M ESPE Express polyvinylsiloxane (PVS), 3M ESPE Permadyne polyether (PE), and Identium (PVSE), impression materials were compared.They were supplied by dispensing from auto-mixing cartridges and were used according to manufacturers’ instructions. Tear strength test and dimensional changes test and contact angle test were evaluated for three elastic impression material.

Wettability test

Wettability assessed by measuring the advancing contacting of liquid on the surface of the set impression material. Controlled (0.1 ml) volume droplet of distilled water was placed onto specimen surface by means of a micropipette and after one minute measure the angle between the surface of the drop and the surface of specimen by dino-litemicroscope was used for imaging the shape of a water drop on the impression material sample surface. The captured image was analyzed by using ImageJ software (ImageJ; USA) to determine the contact angle.12

Tear strength test

Specimens were divided into three groups (n=5). Groups were immediately removed from the mold and loaded in tension until failure using an Instrontesting device (Mensanto,England,Model,WDW-IOOE,No. TC914).

V-notched standard tear strength plastic mold and its riser with the following dimensions was used in this study ( 101.6mm, 19.5mm,2 mm, length, width and thickness respectively) with v-notched region, this was done according to ISO 34-1:2010,The tear strength was calculated using the following formula:

Ts =F/d

Where

Ts: the tear strength (N/mm),

F: the maximum force, in Newton, applied to cause rupture of the specimen,

d : the specimen thickness (mm).

Dimensional accuracy test

Model fabrication

Two block shaped stone models (30mm*20mm*20mm) length, width, height respectively were fabricated. Two holes with a depth of 9 mm were made at 10 mm intervals in each model. These holes will be used to embed the implant analogs. In respect to the first model, the implant holes were prepared with 0–degree angulation, while for the second model the first implant hole was prepared at 0- degree angulation and the second implant hole was prepared at 15- degree angulation.

Implant analogues (Dentium, Seoul, Korea)were inserted in the first hole of each block.Implant analog at 0-degree is usedto serve as the reference point angulation, and another implant analogswas inserted in the second hole at an angulation of 15°degrees. The top of the analog was positioned 1 mm above the model.

Fabrication of the special tray

Impression copings were attached to the analogues. Two sheets of modelling wax (Cavex, Holand) were placed to provide space for the impression material.

A cold cured acrylic material was adapted closely to the wax spacer (Super acrylp®P Plus, Czech Republic).

The marked outline of the tray was trimmed. The access was used to construct the handle following the manufacturer’s instructions. The tray from the master cast was then removed.

The periphery of the tray was trimmed using an acrylic trimming bur to provide mechanical retention for the impression material, hole of 2mm in diameter were drilled at 10 mm intervals. Stops have been provided in the special tray to ensure the uniform thickness of impressions material.Furthermore, the trays were placed in ivomet to minimize the porosities and to obtain better adaptation.

Impression Making

Three impression material were tested,the impression was taken using a closed-tray indirect technique.All impressions for both the parallel assemblies and the divergent assemblies were performed by the same examiner.

In each master model, impression procedure will be repeated for four times according to the type of impression materials used. A total of 12 impressions procedure will be conducted for each individual master model, giving a total of 24 impressions. Material was injected around the impression coping followedbyloading the impression tray with material to seat it on the reference model with gentle finger pressure. Immediately after placing the special tray over the master cast, any excess material was wiped off to verify the complete setting of each tray. The regular set light body impression material were allowed to set as recommended by the manufacturer’s instructions for each one of them. Any remaining access impression material was trimmed .Furthermore, the tray was removed after the material was set completely.

Closed-tray impression copings remaining on the master casts upon removal of the tray after the impression material polymerized. These copings were removed one at a time from the master casts and attached to an implant analog. The impression analogue assembly was inserted into the impression by firmly pushing it into place to full depth. Care was taken to ensure the proper seating of the implant replicas in the impression holes. After 15 minutes impressions were poured with high-strength low-expansion die-hard stone (Zhermack technical, Italy), 100g mixed with 20ml of water. The stone was mixed and poured on a vibrator. After one hour and when the stone has set, casts were separated from the impressions and then trimmed and labeled to prepare for the measurements procedure.

Measurement Protocol

All forty experimental casts were measured and examined for linear dimensional accuracy.  The distance between the impression coping / analogue assembly on the master model and on the study model was measured and compared using a measuring dino-litemicroscope (Dino-lite, Taiwan). It consists of a screen with horizontal and vertical reference lines and was equipped with a light source to project a magnified image of the object onto the screen in the form of a shadow (original magnification ×10),three measurements were made per specimen, and the mean values were computed.Measurements were performed by the same operator to minimize the source of error.

Statistic analysis

One-way ANOVA and Bonferoni test was conducted to analyze study data using IBM SPSs software (Version.23). A P value of > 0.05 was considered statically non-significant (N.S.), ≤ 0.05 was considered significant (S.) and < 0.01 was considered as highly significant (H.S.).

Results

Tear strength test

The mean and standard deviation values for tear strength of the three elastic impression material assessed in this study are displayed in (Table 1)

Among the three experimental groups PE impression material exhibited the higher mean values of tear strength (4.261 N/mm), followed by ID group (4.232N/mm) thenPVS (2.420N/mm) which showed the lowest mean value.

According to ANOVA means of all experimental groups. There were highly significant differences between groups (p < 0.01) (Table 2)

Bonferoni post hoc test was conducted to compare mean values of all groups .There was a statistically significant difference from each other (p<.05), except between PE and ID groups which was non-significant difference (p=0.960) (Table3).

Table 1: Descriptive statistical analysis for tear strength test.

  N Mean Std. Deviation Std. Error 95% Confidence Interval for Mean Minimum Maximum
Lower Bound Upper Bound
PE 7 4.261 1.147 0.433 3.199 5.322 2.69 5.75
ID 7 4.232 0.764 0.288 3.525 4.939 3.47 5.32
PVS 7 2.420 1.208 0.456 1.303 3.537 1.15 4.18

 

Table 2: ANOVA Table for all models included in tear strength test.

Sum of Squares df Mean Square F Sig.
Between Groups 15.571 2 7.786 6.948 0.006 (HS)
Within Groups 20.172 18 1.121
Total 35.743 20

 

Table 3: Multiple comparisonBonferoni test for models included in tear strength test

PE ID 0.028 1.000
PVS 1.840 0.013
ID PE -0.028 1.000
PVS 1.812 0.015
PVS PE -1.840 0.013
ID -1.812 0.015

 

The mean difference is significant when P values< 0.05.

Wettability test

As for the wettability test, the mean descriptive values of the three   impression materials were illustrated in (Table 4). Poly ether group exhibited the lowest mean values of contact angle (39.71000), followed by mean values of contact angle of  ID group  (46.8) ,while the PVS group showed the highest contact angle (50.02).However ANOVA displayed a statistically significant difference was observed between the groups (p<0.05) (Table 5),Bonferoni post hoc testwas conducted for all the results, the ID ,PVS and PE showed  a  statistically significant from each other,  except for  ID and PVS groups which was no significant difference between them( p=0.008) (Table 6).

Table 4: Descriptive statistical analysis for wettability test.

N Mean Std. Deviation 95% Confidence Interval for Mean Minimum Maximum
Lower Bound Upper Bound
ID 10 46.80 3.661 44.180 49.419 41.700 50.100
PVS 10 50.02 0.567 49.614 50.425 49.400 50.900
PE 10 39.71 0.593 39.285 40.134 38.900 40.300

 

Table 5: ANOVA Table for all models included in wettability test.

Sum of Squares Df Mean Square F Sig.
Between Groups 556.442 2 278.221 59.268 0.000
Within Groups 126.745 27 4.694
Total 683.187 29

 

Table 6: Multiple comparison Bonferoni test for models included in wettability test

Mean Difference (I-J) Sig.
ID PVS -3.22 0.008
PE 7.09 0
PVS PE 10.31 0

 

The mean difference is significant when P values< 0.05.

Dimensional accuracy 

Descriptive statistical analysis for parallel implant model after linear measurements of the distance between the heads of the two coping/analogs assembly of the reference model which was equal to (2.004mm). Mean values of the distance between the two parallel assemblies which  were obtained after setting of stone of the working models was (1.902mm, 1.180mm, 1.163mm) for ( ID,VPS ,PE) respectively  (Table 7).

Regarding the 150 angulation between the heads of the two coping/analogs assembly the of the second reference model the distance between the two heads was equal to (2.356mm). Mean values of the distance between the heads of  15 0 angled assemblies which were  obtained after setting of stone of the working models was (2.246mm, 1.945mm, 1.163mm) for ( ID,VPS ,PE) respectively . Overall results are provided in (Table 7).

Table 7: Descriptive statistical analysis for dimensional stability test.

    N Mean Percentage changes from the control Std. Deviation 95% Confidence Interval for Mean   Minimum Maximum
            Lower Bound Upper Bound    
Straight Control 10 2.004 0 0.0106 1.996 2.011 1.98 2.023
ID 10 1.902 0.05 0.0811 1.844 1.96 1.705 1.994
VPS 10 1.18 0.41 0.0052 1.176 1.183 1.175 1.185
PE 10 1.163 0.42 0.0151 1.152 1.174 1.132 1.175
Angled Control 10 2.356 0 0.0497 2.321 2.392 2.272 2.424
ID 10 2.246 0.046 0.0714 2.195 2.297 2.093 2.303
VPS 10 1.945 0.17 0.037 1.918 1.971 1.9 2.009
PE 10 1.918 0.18 0.0044 1.914 1.921 1.91 1.923

 

Analysis of the data by ANOVA Table demonstrated significant differences in linear distance measurements among groups for both the parallel and angled assemblies of the implant models (p < 0.01) (Table 8).

Table 8: ANOVA Table for all models included in dimensional stability test.

Sum of Squares Df Mean Square F Sig.
Straight Between Groups 6.159 3 2.053 1181.886 0.000
Within Groups .063 36 .002
Total 6.222 39
Angled Between Groups 1.432 3 .477 212.598 0.000
Within Groups .081 36 .002
Total 1.513 39

 

Bonferroni post hoc test revealed significant differences in linear distance measurements between the two reference models the parallel one and the angulated (group R1, R2) and their belonged experimental groups (ID, PVS, PE). However, there was a significant difference among the three groups for both conditions, on the other hand there was no significant difference when readings were observed between (PE and PVS) (Table 9).

Table 9: Multiple comparison deferential statistical analysis for dimensional stability test.

Dependent Variable     Mean Difference (I-J) Sig.
Straight R1 ID 0.101 0
PVS 0.824 0
PE 0.84 0
ID PVS 0.722 0
PE 0.738 0
PVS PE 0.016 1
Angled R2 ID 0.11 0
PVS 0.411 0
PE 0.438 0
ID PVS 0.301 0
PE 0.328 0
PVS PE 0.027 1

 

The mean difference is significant when P values< 0.05.

Discussion

The null hypothesis was rejected because there were significant difference between the three materials regarding the results of tear strength, wettability and accuracy.From the standpoint of clinical application, the ideal impression material should exhibit high tear strength with maximum energy absorption and minimal distortion.13 The result revealed that the tear strengthof PE and IDENTIUM were significantly higher than PVS, and  show no significant difference between each other, this came in contrary with Hondrum et al.in 199414 who concluded that there were no significantly different between PVS and PE. Also, the results were in disagreement with Lawson et al.in 200815 who found that PVS showed higher tear strength than PE and PVSE. The reason for this difference may be due to different test method, condition, setting time.

According to the analysis of the mean values of the contact angle obtained from the wettability test, the PE showed the highest degree of wettability followed by ID and PVS impression material.Bonferoni test demonstrated that there was a significant difference between PE and ID group, and also PE and PVS impression materials,however there was no significant difference between ID and PVS.

Higher values of contact angles results in more hydrophobicity, and low values of contact angles results in more hydrophilisity.9 The results of this study was comparable to  the study of Michalakis et al ,2007,16 who examined and  compared the hydrophilicity of six elastomeric impression materials before and after setting  and concluded that PE exhibited  the most hydrophilic among all evaluated material.

The recently produced hybrid structure PVES was a composite of two material polyether and polyvinylesiloxane which get benefit from the properties of both material,it’s a chemical structure contains polyether which reputed for it hydrophilisity, wettability and precise castability. It is suggested that the addition of polyether to PVS can increase the hydrophilisity of PVS and castability without the need for adding surfactants before impression, pouring.This phenomenon can be explained as the following:  the mixture of high molecular weight of polyether chains form the backbone frames, and that the smaller PVS molecules attach to the PE backbone. The existence of functional groups of VSE can provide similar hydrophilic characteristics to PE.17

PE impression material is claimed by its manufacturer as more hydrophilic because of its functional groups [carbonyl (C = O) and ether (C-O-C)]. These oxygen group has more affinity to water.18,19 These polarized groups can be attracted and interact with water molecules; this interaction facilitates the contact between impression materials and moist oral tissues [Van Krevelen, 1997].20Conventional PVS behaves hydrophobically because it does not contains any polarized groups. A material exhibiting contact angle of greater than 90° is an indication of poor wetting, which means that the material exhibits hydrophobicity, while a material exhibiting contact angle of less than 90° are an indication of better wetting, which means that it exhibits hydrophilicity.21

Additionally, our results was comparable to the finding of Shetaetal, in 201720 who concluded that PE and VPES exhibited the higher wettability when the compared with two other group of PVS.

Obviously from the obtained results about the PE and IDENTIUM hydrophilicity, it is inherent in nature for both without the need of adding extrinsic material.

Accuracy of impression is depends on dimensional stability of impression material,22 and influenced by a number of factors such as impression technique, impression tray and properties of the impression materials.23 An accurate impression is an important step in processing and final fitting of dental prosthesis.24

Fixture level impression is crucial especially in cases where angulation between abutments and vertical spaces are difficult to be assessed intra-orally.25

Two impression coping/analogue assembly were placed in each reference model. The first reference model was fabricated with two parallel impression coping/analogue assembly, while the second reference model was fabricated with 15-degree angulations between the fixed assemblies, in order to simulate common clinical situations that may necessitate placement of angulated implants. Furthermore, in contrary of most of previous studies, the implants in this study were tilted to the distal side.26

With regard to the results compared with the first reference models,measurements(R1),there was a significant difference in the linear distance measurements between the two parallel impression coping/analogue assembly of the working models obtained  by using (Identium, 3M ESPE ExpressPVS , 3M ESPE Permadyne polyether)(2.24644mm , 1.94535mm, 1.16380 mm) respectivelyas an  impression material  and measurement of their  reference model(2.35670mm). Also the results displayed a significant difference among all the experimental group, with no significant difference observed between polyether and PVS group, this came in agreement with other studies  that there was no difference between PVS and PE in multi- implant impressionwith Vojadnietal,2015.27

Some studies suggested that PE and PVS perform an accurate results minimum amount of distortion and adequate rigidity.28

Other study ascertained the superiority of PVS in comparison with PE.29 Some studied indicated the superiority of PE in parallel condition compared to angulate one.30

In the present study the PE showed some degree of distortion in both parallel and angulated condition. Furthermore, no superiority in comparison between parallel and non-parallel conditions also when it was compared with ID and PVS impression materials.

Attributed lower rigidity of PVS, is considered as alternative. It can be used more safely in partially edentulous particularly in nonparallel situations or in cases with sever undercuts area or multiunit implants and in cases with subgingival implant placement.31 According to Del’acqua et al. in 200832 ether should be the material of choice to achieve a more accurate orientation of implant analogues in laboratory master casts. The author also stated that the material rigidity prevents displacement of impression copings within the impression material.

However, Enklinget al,in 2012 concluded that there was no significant difference between polyether and PVS.Identium indicated superiority about subjective assessment of the dentist (handling, taste, precision detail of impression) and the dental technician (the ease of removing plaster model from the mold.33

The results of dimensional measurements in non-parallel conditions show that the polyvinyl siloxane isthebest choice, followed by vinyl siloxanether and polyether.

The results of this study were contradictory with Vojdaniin 2015, the ID showed a significant difference with PSV and PE. Though,the study detected a significant degree of distortion of ID impression material compared with measurement of R1 and R2, mean values of the working models,measurements were more accurate than PE and PVS mean values.34

However it is noteworthy that statistical analysis of the measurements’ indicated that distortion in all of the impression materials in both parallel and angled condition but consistently.

Some studies claimed that the use of two or three angled implants found to express no significant differences between the angled and parallel implants in terms of misfit.35 Other studies concluded that degree of error is inevitable in all the impression transfer protocols studied.36

The working models produced by all 3 experimental impression materials were distortedin both conditions (angled and parallel implant).The measurements detected a reduction in mean values between the reference model and the working cast,the distance between the impression and coping assemblies which is extrapolated to mesio-distal direction in clinical situation. The reduction may be attributed to the shrinkage of impression material toward the center of the mass.This finding was in coordinate with Ceyhan etal,in 2003 who found that the distortion of the impression is a concerninherent, in a three-dimensional way, in all ofthe procedures involved in the indirect dentalrestorations.37

Other studies attributed the distortion to the possible inaccuracy of fixture-level impressionsdue to difficultiesin repositioning the impression coping correctly in elastic material.38,39

ADA specification No. 19 described the criteria that elastomeric impression materials should not display more than 0.5% dimensional change after 24 h of polymerization of the material,40all materials used in this study were within the accepted stander .

References

  1.  Arcangelo D.C, Zarow M,  Angelis D.F, Vadini M, Paolantonio M, Giannoni M, Amario D.M. Five-year retrospective clinical study of indirect composite restorations luted with a light-cured composite in posterior teeth. Clin Oral Investig. 2014;18(2):615-24.
    CrossRef
  2. arcangelo D.C, Angelis D.F, Vadini M, Amario D.M. Clinical evaluation on porcelain laminate veneers bonded with light-cured composite: results up to 7 years. Clinical oral investigations. 2012;16(4):1071-9.
    CrossRef
  3. Al-Kaisy N. A Survey of Prosthodontics Techniques Applied by Dental Practitioners in Sulaimani City. J Bagh Coll Dentistry. 2016;28(3):22-29.
    CrossRef
  4. Herfort T.W, Gerberich W.W, Macosko C.W, Good kind R.J.  Tear strength of elastomeric impression materials. J Prosthet Dent. 1978 Jan; 39(1):59-62. C.
    CrossRef
  5. Magalhães C.S, Moreira A.N.  Dimensional stability of elastomeric impression materials: a critical review of the literature. Eur J Prosthodont. Restor Dent. 2011 Dec;19(4):163-6.
  6. Reddy G.V, Reddy N.S, Itttigi J, Jagadeesh K.N. A comparative study to determine the wettability and castability of different elastomeric impression materials. J Contemp Dent Pract. 2012 May 1;13(3):356-63.
  7. Petrie C.S, Walker M.P, Mahony O.A.M and Spencer P.  Dimensional accuracy and surface detail reproduction of two hydrophilic vinyl polysiloxane impression materials tested under dry, moist, and wet conditions. J Prosthet Dent. 2003;90:365–372.
    CrossRef
  8. Lu H, Nguyen B, Powers J.M.  Mechanical properties of 3 hydrophilic addition silicone and polyether elastomeric impression materials. J Prosthet Dent. 2004 Aug;92(2):151-4.
    CrossRef
  9. Kess R.S, Combe E.C, Sparks B.S.  Effect of surface treatments on the wettability of vinyl polysiloxane impression materials. J Prosthet Dent. 2000 Jul;84(1):98-102.
    CrossRef
  10. Nassar U, Oko A, Adeeb S, El-Rich M , Flores-Mir C.  An in vitro study on the dimensional stability of a vinyl polyether silicone impression material over a prolonged storage period. J Prosthet Dent. 2013 Mar;109(3):172-8.
    CrossRef
  11. Pandita A, Jain T, Yadav N.S, Feroz S.M, Pradeep, Diwedi A. Evaluation and comparison of dimensional accuracy of newly introduced elastomeric impression material using 3D laser scanners: an in vitro study. J Contemp Dent Pract. 2013 Mar 1;14(2):265-8.
    CrossRef
  12. Zgura L, Beica T, Mitrofan I.L, Mateias C.G, Pirvu D, Patrascu I. Assessment of the impression materials by investigation of the hydrophili city. Dig J Nanomater Bios 2010;3:749-55.
  13. Lu H, Nguyen B, Powers J.M.  Mechanical properties of 3 hydrophilic addition silicone and polyether elastomeric impression materials. J Prosthet Dent. 2004 Aug;92(2):151-4.
    CrossRef
  14. Hondrum S.O.  Tear and energy properties of three impression materials. Int .J Prosthodont. 1994;7:517–521.
  15. Lawson N.C, Burgess J.O, Litaker M.S.  Tear strength of five elastomeric impression materials at two setting times and two tearing rates. J Esthet Restor Dent. 2008;20:186–194.
    CrossRef
  16. Michalakis K.X, Bakopoulou A, Hirayama H, Garefis D.P, Garefis P.D. Pre- and post-set hydrophili city of elastomeric impression materials. J Prosthodont. 2007;16:238–248.
    CrossRef
  17. Erkut S, Can G.  Effects of glow-discharge and surfactant treatments on the wettability of vinyl polysiloxane impression material. J Prosthet Dent. 2005 Apr;93(4):356-63.
    CrossRef
  18. Craig R.G, Powers J.M.  Restorative dental materials. 11th St Louis: Elsevier. 2002;330–431.
  19. Mai S.S, El-Shorbagy Z.A, Karim A.U.M, Abd-Alla.  Saeed Laboratory comparative study of wettability, dimensional changes, flexibility and tear resistance of two recent elastomeric impression materials Year 2017;14(2):89-95.
  20.  Krevelen V.D.W.  Properties of polymer: their correlation with chemical structure their numerical estimation and prediction from additive group. 3rd ed. Amsterdam: Elsevier. 1997.
    CrossRef
  21. Chai J.Y and Yeung T.C.  Wettability of no naqueous elastomeric impression materials. Int J Prosthodont. 1991 Nov-Dec;4(6):555-60.
  22. Markovic D, Puskar T, Hadzistevic M, Potran M, Blazic L, Hodolic J.  The dimensional stability of elastomeric dental impression materials. Contemp Mater. 2012; III-1:105–10.
    CrossRef
  23. Stober T, Johnson G.H, Schmitter M.  Accuracy of the newly formulated vinyl siloxane ether elastomeric impression materials. J Prosthet Dent. 2010;103(4):228–39.
    CrossRef
  24. Piwowarczyk A, Ottl P, Buchler A, Lauer H.C, Hoffmann A.  Invitro study of dimensional accuracy of selected materials for monophase elastic impression making. Int J Prosthodont. 2002;15(2):168–74.
  25. Choi J.H, Lim Y.J, Yim S.H, Kim C.W.  Evaluation of the accuracy of implant-level impression techniques for internal-connection implant prostheses in parallel and divergent models. Int J Oral Maxillofac Implants. 2007; 22:761–8.
  26. Geramipanah F, Sahebi M, Davari M, Hajimahmoudi M, Rakhshan V. Effects of impression levels and trays on the accuracy of impressions taken from angulated implants. Clin Oral Implants Res. 2015;26(9):1098–10.
    CrossRef
  27. Carr A.B.  Comparison of impression techniques for a two-implant 15-degree divergent model. Int J Oral Maxillofac Implants. 1992;7:468–75.
  28. Daoudi M.F, Setchell D.J, Searson L.J.  A laboratory investigation of the accuracy of two impression techniques for single-tooth implants. Int J Prosthodont. 2001;14:152–8.
  29. Liou A.D, Nicholls J.I, Yuodelis R.A, Brudvik J.S.  Accuracy of replacing three tapered transfer impression copings in two elastomeric impression materials. Int J Prosthodont. 1993;6:377–83.
  30. Choi J.H, Lim Y.J, Yim S.H & Kim C.W.  Evaluation of the accuracy of implant level impression techniques for internal-connection implant prostheses in parallel and divergent models. Int J Oral Maxill of ac Implants. 2007 Sep-Oct; 22(5):761-8.
  31. Lu H, Nguyen B, Powers J.M.  Mechanical properties of 3 hydrophilic addition silicone and polyether elastomeric impression materials. J Prosthet Dent. 2004;92:151–4.
    CrossRef
  32. Del’Acqua M.A, Arioli-Filho J.N, Compagnoni M.A, MolloFde A. Jr Accuracy of impression and pouring techniques for an implant-supported prosthesis. Int J Oral Maxill of ac Implants. 2008;23:226–36.
  33. Enkling N, Bayer S, Jöhren P, Mericske-Stern R.  A new impression material. Clinical study of implant impressions with vinylsiloxanether versus polyether materials. Clin Implant Dent Relat Res. 2012 Mar;14(1):144-51.
    CrossRef
  34.  Vojdani M, Torabi K, Ansarifard E. Accuracy of different impression materials in parallel and nonparallel implants. Dent Res J (Isfahan). 2015 Jul-Aug;12(4):315–322.
    CrossRef
  35. Conrad H.J, Pesun I.J, DeLong R, Hodges J.S.  Accuracy of two impression techniques with angulated implants. J Prosthet Dent. 2007 Jun;97(6):349-56.
    CrossRef
  36. Geramipanah F, Sahebi M, Davari M, Hajimahmoudi M, Rakhshan V. Effects of impression levels and trays on the accuracy of impressions taken from angulated implants. Clin Oral Implants Res. 2015;26(9):1098–10.
    CrossRef
  37. Ceyhan J.A, Johnson G.H, Lepe X.  The effect of tray selection, viscosity of impression material, and sequence of pour on the accuracy of dies made from dual-arch impressions. J Prosthet Dent. 2003;90:143–149.
    CrossRef
  38. Cehreli M.C & Akca K.  Impression techniques and misfit-induced strains on implant-supported superstructures: an in vitro study. Int J Periodontics Restorative Dent. 2006;26:379–385.
  39. Al-Azawi R. A, Al-Naqash W.A.  The Effect of Silver-Zinc Zeolite Incorporation on Some Properties of Condensation Silicone Impression Material. J Bagh. Coll Dent. 2016;28(4):22-27.
  40. American National Standard/American Dental Association. Specification no. 19 for non-aqueous, elastomeric dental impressions. J Am Dent Assoc. 1977;94:733–741.
    CrossRef
Share Button
(Visited 2,403 times, 1 visits today)

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.